Product Details:
|
Wieght: | 1.65kg | Product Name: | Syringe Pump |
---|---|---|---|
Brand Name: | RAINBOW | MOQ: | 1 |
OEM: | YES | ||
Highlight: | Safe Portable Syringe Pump,Reliable Portable Syringe Pump |
Technology Parameters of Syringe Pump
Main features |
Reliable and Safe: |
Advanced safety design able to stop infusion and alarm even if the CPU damage,the medical staff can use it safely. |
Applicable syringe |
5ml,10ml,20ml,30ml,50/60ml |
The maximum flow rate range |
5ml syringe: 0.1ml/h~200ml/h |
10ml syringe: 1ml/h~300ml/h |
20ml syringe: 1ml/h~400ml/h |
30ml syringe: 1ml/h~600ml/h |
50/60ml syringe: 1ml/h~1200ml/h |
Precision |
Better than ±2%; |
(but 5ml syringe and <1ml/h: Better than ±8%) |
Parameter range |
1, Flow-rate: 0.1~1200.0 ml/h, 0.1ml/h step; |
2, Dose: 0.1~999.9 mg, 0.1mg step; |
3, mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step; |
4, Body-weight: 1~300 kg, 1kg step; |
5, KVO rate: 0.1~5.0 ml/h, 0.1 ml/h step; |
6, Bolus rate: 0.1~1200.0 ml/h, 0.1 ml/h step; |
7, Bolus volume: 0.1~99.9 ml/h, 0.1ml step |
Alarm |
1, Occlusion,0~175KPa, 8 grades of sensitivity; |
2, Pressure sensor is abnormal; |
3, Push rod fall off; |
4, The drug running out; |
5, The injection will finish; |
6, The motor rotation fault; |
7, Overtime pause; |
8, Low battery; |
9, Battery failure; |
10, Battery temperature is too high; |
11, Mains missing tips; |
12, MCU fault, and stop injection. |
Safety classification |
Class I; Internal power supply, CF type, IPX4 equipment |
Power supply |
AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery |
Power consumption:20VA |
Working time of battery: 10h@5ml/h |
Weight and size |
37cm(w)*24cm(h)*21cm(t); about 2.1kg |
Optional accessories |
Wireless communicating module. |
Chinese Certificate |
yes |