Product Details:
|
Weight: | 1.20kg | Brand Name: | RAINBOW |
---|---|---|---|
Highlight: | Transfused Blood Hospital Infusion Pump,ISO13485 Hospital Infusion Pump,Peristaltic Infusion Pump In ICU |
Main features: |
Reliable and Safe: |
1, Additional set of bubble alarm design(double bubble alarm) ensure more secure infusion; |
2, Advanced safety design able to stop infusion and alarm even if the CPU damage,the medical staff can ues it safely; |
3.multipurpose: |
It can be transfused, transfused blood and transported nutrient solution. |
Pumping way: |
Semi-squeeze finger-type peristaltic pump |
IV-sets: |
Special or quality PVC/TPE infusion sets with an external diameter of 3.4mm~4.5mm |
User options: |
No |
Infusion modes: |
4 infusion modes: |
1, Flow-rate: ml/h; |
2, Time-amount: ml/h; |
3, Drop-rate: d/min; |
4, Body-weight: mg/kg/h |
Parameter range |
1, Flow-rate: 1.0~1200.0 ml/h, 0.1ml/h step; |
2, Time: 1s~99h59m59s, 1s step; |
3, Drop-rate: 1~333 d/min, 1 d/min step; |
4, Dose: 0.1~999.9 mg, 0.1mg step; |
5, mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step; |
6, Body-weight: 1~300 kg, 1kg step; |
7, KVO rate: 1.0~5.0 ml/h, 0.1 ml/h step; |
8, Bolus rate: 50.0~800.0 ml/h, 0.1 ml/h step; |
9, Bolus volume: 0.1~99.9 ml/h, 0.1ml step |
Precision |
Better than ±5% (using a calibrated or a special IV-set) |
Alarm |
1, Air in tube, 5 grades of sensitivity (double bubble alarm); |
2, Bubble sensor fault; |
3.1, Upper-end occlusion, -20~ -60 kPa,8 grades of sensitivity; |
3.2, Upper-end base-pressure fault; |
4.1, Lower-end occlusion, 20~ 140 kPa,8 grades of sensitivity; |
4.2, Lower-end base-pressure fault; |
5.1, Pump rotation fault; |
5.2, Inverse-infusion; |
6, Abnormal door-opening; |
7, Low battery power, battery fault; |
8, Overtime pause; |
9, Completed infusion; |
10, AC power fault; |
11, MCU fault, and stop infusion; |
12, Battery temperature is too high. |
Data function: |
1, Memory of calibration: Records the calibration data of tube automatically for application; |
2, Memory of parameter: Records the latest infusion parameter for next operation; |
3, Historical records: records latest 1000 infusion events. |
(what/when/why,flow-rate, etc.) |
Safety classification: |
Class I; Internal power supply, CF type, IPX4 equipment |
Power supply: |
AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery |
Power consumption:20VA |
Working time of battery: 7h@25ml/h |
Weight and size: |
130mm(w)*206mm(h)*118mm(t); about 1.6kg |
Optional accessories: |
1, Warmer; |
2, Wireless communicating module. |
Chinese Certificate: |
yes |
Contact Person: Lily